An Electrochemical DNA Sensor for Ultrasensitive Detection of Aspirin Resistance Gene Based on Signal Amplification Strategy of Ru(NH3)63+/MPA/Cys Redox Cycling
1Department
of Orthopedics, affiliated Hospital of Southwest Medical University, Luzhou,
China
2Department of rehabilitation Medicine, affiliated Hospital of Southwest Medical University, Luzhou, China
*Corresponding author: Xiaobo Lu, Department of Orthopedics, PROOF affiliated Hospital of Southwest Medical University, No. 20 of Taiping Street in the Jiangyang district, Luzou city, Sichuan province, P. R. China. Tel: +86-8313165451; Fax: +86-8313165451 Email: wyx20170820@163.com
Received
Date: 21 May, 2018; Accepted Date: 04 June, 2018; Published Date: 11 June, 2018
Citation: Wan Y, Xu L, Zhuo N, Lu X (2018) An Electrochemical DNA Sensor for Ultrasensitive Detection of Aspirin Resistance Gene Based on Signal Amplification Strategy of Ru(NH3)63+/MPA/Cys Redox Cycling. Biosens Bioelectron Open Acc: BBOA-136. DOI: 10.29011/ 2577-2260. 100036.
1. Abstract
Aspirin resistance, the phenomenon of patient showing no response to aspirin, is characterized by a predominant mutation in the COX-1 gene. Accordingly, patients are routinely screened for mutations in this gene to determine whether they can benefit from aspirin treatment. In this work, we reported a simple and ultrasensitive electrochemical method for COX-1 C50T detection for the first time based on Electrochemical-Chemical-Chemical (ECC) redox cycling. In the procedure of assay, Ru(NH3)63+ is used as the electron acceptor that is electrostatically attracted to Peptide Nucleic Acid (PNA) modified electrodes and new chemical regents of Mercaptopropionic Acid (MPA) and cysteamine were introduced, which permits Ru(NH3)63+ to be regenerated for multiple redox cycles. Different Pulse Voltammetry (DPV) was applied to record the electrochemical signals, which increased linearly with the target DNA. Under optimal conditions, the DNA sensors showed a wide linear relationship, from 10 fM to 1 nM, with detection limits of 2.8 fM (S/N = 3). This strategy paves a new avenue for the determination of aspirin resistance gene in patient serum.
2.
Keywords:
Aspirin Resistance; COX-1; DNA Sensor; Electrochemical-Chemical-Chemical
1. Introduction
After more than a century, aspirin is one of the most commonly used drugs in western medicine for the prevention of cardiovascular diseases [1]. Despite adequate aspirin therapy, 6.2% to 57% patients still experience cardiovascular events, the phenomenon named as Aspirin Resistance (AR) [2]. Researchers have reported that genetic variability in Cyclooxygenase-1 (COX-1) C50T affect aspirin response with cardiovascular disease and could be used as specific genetic marker of AR [3,4]. Therefore, it is crucial to prescreen patients for their genotypes before prescribing aspirin to facilitate target therapy and decreasing the side effects associated with prolonged administration.
Various methods have been developed for the detection of the COX-1 mutation at the DNA level. Among them is the Polymerase Chain Reaction (PCR) or DNA sequencing [5,6]. DNA sequencing is an excellent approach for research studies that seek to profile large regions of DNA, but its implementation is prohibitively expensive for routine clinical use and the slow turnaround time (2-3 weeks) is not an ideal for optimal treatment outcomes.6 Also, PCR is prone to interference from the components of biological samples and requires purification of the nucleic acids from samples. Therefore, a method that is simpler, and able to detect rare mutations directly in serum, is required urgently.
In recent years, a versatile platform for markers detection has been developed based on DNA sensors, owing to their low cost and potential for high levels of multiplexing and sensitivity [7]. This type of testing approach has been applied successfully to the analysis of a subset of antibiotic resistance [8], but the feasibility of analysing resistance gene for anticoagulant drug in clinical samples has not been established. Here we explored the feasibility of applying this technology to direct analysis of the aspirin resistance gene COX-1 from patient serum samples. To minimize the background current and increase the signal-to-noise ratio, a neutral charge biotinylated Peptide Nucleic Acid (PNA) probe complementary to the target sequence was selected to modify the electrode surface [9,10]. For accomplishing the low abundance sequence detection, streptavidin was further introduced during DNA sensor fabrication. Because it can provide more active sites for the immobilization PNA probe [11,12]. In order to immobilization more streptavidin, reduced Graphene Oxide–Tetraethylene Pentamine (rGO–TEPA) and Gold Porous Nanoparticles (Au PNPs) were integrated [13-15].
Electrochemical-Chemical-Chemical (ECC) redox cycling can be used to amplify signals produced at DNA modified electrode surfaces to increase the sensitivity and accuracy of a detection assay [16-18]. In the procedure of assay, Ru(NH3)63+ acts as a redox label and binds to DNA nonspecifically through electrostatic interactions with the phosphate backbone, and therefore, the electrochemical reduction of this species yields a signal that reports on the increase of negatively charged groups at the electrode surface upon hybridization of a target sequence [19]. In addition, to overcome the interfering redox reactions near the potential of interest and amplify signals, new chemical regents of MPA and Cys were introduced, which permits Ru(III) to be regenerated for multiple redox cycles [20]. In recent years, although promising research has produced DNA sensors using ECC redox cycle reporter system to detect mutated nucleotides in samples, to the best of our knowledge, this is the first reported use of DNA sensors for the detection COX-1 gene, a biomarker implicated aspirin resistance.
2. Expérimental
2.1. Chemicals and Materials
Reduced Graphene Oxide – Tetraethylene Pentamin (rGO-TEPA) was provided by Nanjing XFNANO Materials TECH Co., Ltd. (China). Tris-(2-Carboxyethyl)-Phosphine Hydrochloride (TCEP), Gold (Ш) Chloride Trihydrate (HAuCl4·3H2O), streptavidin, 6-Mercaptohexanol (MCH), Ru(NH3)6Cl3, 3-Mercaptopropionoic Acid (MPA), Cysteamine (Cys), were obtained from Sigma-Aldrich (St. Louis, USA, www. Sigma-Aldrich. com). Potassium Ferricyanide (K3Fe(CN)6) and Potassium Ferrocyanide (K4Fe(CN)6) were purchased from Beijing Chemical Reagents Company (Beijing, China). Clinical serum samples were obtained from a local hospital and stored at 4°C. Biotinylated Peptide Nucleic Acid (PNA) probes were synthesized and purified by Shanghai RuiDi Biological Technology Co., Ltd. The sequences of PNA probes are as follows: 5’-Bio-Cys-O-ATC-ATC-CTG-ACT-GGC-ATC-CGG-3’, where Bio represents biotin, O represents an ethylene glycol linker and Cys represents the amino acid cysteine. Other oligonucleotides were also synthesized and purified by Shanghai RuiDi Biological Technology Co., Ltd. with their sequences illustrated in Table S1.
The
buffer solutions involved in this study were as follow: 1 × TE buffer (10 mM Trishydroxymethylaminomethane
hydrochloride (Tris-HCl), and 1.0 mM Ethylenediaminetetraacetic Acid (EDTA), pH 8.0),
which was used to dissolve all oligonucleotides. Phosphate Buffered Solution (PBS)
(pH 7.4, 0.1 M) was prepared with NaH2PO4
and Na2HPO4. Tris buffer solution (50 mM, pH 9)
was obtained from Sigma-Aldrich (St. Louis, USA, www. Sigma-Aldrich.
com). All other reagents were of
analytical reagent grade
and used without further purification. Ultrapure water (>18.2 ΩM) obtained from a Millipore Mill-Q purification system
was used throughout the experiment.
All electrochemical experiments were carried out on a CHI660E electrochemical workstation (Chenhua Instruments Co., Shanghai, China). Scanning Electron Microscopy (SEM) images were obtained by using a Hitachi S-3000N (Hitachi Limited, Japan). Field Emission Scanning Electron Microscopy (FE-SEM) image was conducted using a Hitachi S4800 (Hitachi Limited, Japan). Transmission Electron Microscopy (TEM) investigations were performed on a Hitachi Limited, Japan). A conventional three-electrode system was used for all electrochemical measurements: a Ag/AgCl reference electrode and a Pt counter electrode, and the modified Glassy Carbon Electrode (GCE, 4-mm in diameter) as the working electrode. A one-compartment cell fitted with a Luggin capillary was used to separate the working compartment from the reference compartment. Quantitative real-time Polymerase Chain Reaction (qRT-PCR) experiments were performed on a Light Cycler Nano Real Time PCR Instrument (Roche, UK).
2.3. Preparation of Porous Au Nanoparticles
Porous Au NPs were synthesized referencing to the literature [21,22]. Briefly, 1 M KOH solution (0.05 mL), 10 mM fresh ascorbic acid (3 mL) aqueous solution, and 2 mL H2O were sequentially injected into a conical flask. Then 2 mM Gold (Ш) chloride trihydrate solution (4 mL) was added. The mixture was blended for about 1min and then located in 60℃ thermostated container for 1 h. The resulting Au PNPs were centrifuged and washed with absolute ethanol and deionized-distilled water for thrice and then dispersed in 2 mL Milli-Q water.
2.4. Construction of DNA Sensors
The
fabrication procedure of the electrochemical DNA biosensor is illustrated in Scheme 1. Prior to use, GCE was polished repeatedly
using 0.3- and 0.05-µm alumina slurries
followed by thorough rinsing with Milli-Q water. After successive sonication in baths of Milli-Q water, absolute alcohol and
Milli-Q water again, the electrode was dried at room temperature. After that,
10 µL of rGO-TEPA (2 mg mL-1) was cast onto the electrode surface
and dried in the air. Subsequently,
the electrode was immersed in porous Au nanoparticles solution and connected to
CHI660E workstation using a three-electrode set-up with a Ag/AgCl reference
electrode and a Pt counter electrode.
Au PNPs were electrochemically deposited onto the rGO-TEPA-modified GCE surface under a constant potential of -0.2 V for 30 s. After washing with Milli-Q water, 10 µL of a streptavidin solution (100 ng mL-1) was dropped onto the modified electrode surface through the strong effect of Au-NH2 bonding to immobilize streptavidin. The electrode was kept at 4°C for 16 h, as streptavidin can retain its bioactivity for a long period of time at this temperature. After the reaction, the electrodes were washed with Milli-Q water to remove loosely bound chemicals and dried in nitrogen. After that, the device was covered with a EP lid.
2.5. Functionalization of DNA Sensors
An aqueous solution containing 1 µM of probe was mixed with 10 µM of aqueous Tris(2-carboxyethyl) phosphine hydrochloride solution and the mixture was left for overnight to cleave disulphide bonds. After that, 10 µL of probe solution was pipetted onto the sensors and incubated for 10 h in a dark humidity chamber at room temperature. Through the specific recognition between streptavidin and biotin, the biotinylated peptide nucleic acid probe was immobilized on the electrode surface for hybridization with target DNA. The sensors were then washed thrice for 3 min with 0.1 × PBS at room temperature. The sensors were then treated with 1 mM of 6-Mercaptohexanol (MCH) for an hour at room temperature for back filling. After washing, the sensors were challenged with different concentration of targets for 60 min at room temperature. After hybridization, the sensors were washed thrice for 3 min with 0.1 × PBS at room temperature and the electrochemical scans were acquired.
2.6. Electrochemical detection of COX-1 DNA
All electrochemical experiments were carried out using CHI660E electrochemical workstation with a three-electrode system featuring a Ag/AgCl reference electrode and a platinum wire auxiliary electrode. The accumulation of Ru(NH3)63+ was measured by immersing PNA/DNA-modified electrodes in 10 µM Ru(NH3)6Cl3 for 60 s, an amount of time where the cation would have equilibrated. By transferring to a Tris buffer solution (50 mM, pH 9) containing 10 µM Ru(NH3)6Cl3, 0.5 mM 3-Mercaptopropionoic Acid (MPA) and 0.5 mM Cysteamine (Cys), Differential Pulse Voltammetry (DPV) signals were obtained with the potential range was from 0 V to 0.5 V, pulse amplitude of 50 mV, pulse width of 50 ms, and a pulse period of 100 ms. Signal changes that correspond to specific targets were calculated with background-subtracted currents: change in current (∆I) = Iafter - Ibefore, where Iafter = current after target binding and Ibefore = current before target binding.
3. Results and discussion
3.1. Principle of this Method
We use a multi-pronged strategy to minimize the current in the absence of target nucleic acid. The sensors are functionalized with biotinylated Peptide Nucleic Acid (PNA) probes complementary to the target sequence. PNA is a synthetic nucleic acid analogue that has a neutral charge. This neutral charge minimizes the background current and increases the signal-to-noise ratio. Our new assay employs Ru(NH3)63+, Mercaptopropionic Acid (MPA) and cysteamine. The Ru(NH3)63+ electron acceptor complex is positively charged and binds to the sensors at levels that corresponding to the amount of negatively charged nucleic acid. On the application of a potential at 250 mV, Ru(NH3)63+ is oxidized to Ru(NH3)64+, producing a measurable current that reports on the presence of the target DNA. However, the limited concentration of Ru(NH3)63+ localized at the bound target nucleic acids yields a small current. To provide maximal sensitivity for the detection of DNA hybridization, we introduced MPA, the MPA is neutral charge, and so it does not bind to the sensor, but presents in solution to chemically reduce Ru(NH3)64+ back to Ru(NH3)63+, allowing for multiple turnovers of Ru(NH3)63+, which generates a high electrocatalytic current. This signal is further amplified by cysteamine, another reduction agent, which is chemically oxidized to cystamine by reducing the oxidized form of MPA (R-S-S-R) back to its reduced form (R-SH).
3.2. Characteristics of the Modified Electrodes
The
stepwise DNA sensors fabrication processes were investigated by SEM and FE-SEM.
(Figure 1A, 1a) display the typical SEM and
FE-SEM images of rGO-TEPA. It
was confirmed that rGO-TEPA has
a wrinkled paper-like structure. SEM of the rGO-TEPA/Au PNPs is shown in (Figure 1B). As seen, spherical nanoparticles were
coated on the rGO-TEPA surface.
(Figure 1b) displays a typical TEM image of Au
PNPs, which clearly reveals Au PNPs has porous spherical morphology and the
average size of the nanoparticles was approximately 30 nm. Upon immobilization
of streptavidin, proteins are trapped on the surface of the rGO-TEPA/Au PNPs, and a blurry
surface image was observed, distinct from that shown in (Figure 1B). This result is indicative of successful streptavidin
bonding on the surface of the rGO-TEPA/Au
PNPs. Afterwards, with the immobilization of the biotinylated PNA
probe onto the rGO-TEPA/Au PNPs/streptavidin
film, the modified nanostructure exhibited a crystal shape, as shown in (Figure 1D). This crystal shape is distinct from that
shown in (Figure 1C), indicating that the
biotinylated PNA probe was successfully introduced via specific recognition
between streptavidin and biotin.
3.3. Electrochemical Behavior of DNA Sensors
The electrochemical characteristics of different nanomaterials were investigated. It is clearly seen in (Figure 2A) that the electrode when modified by rGO-TEPA as compared with glassy-carbon electrode (curve a), and the electrode modified by Graphene Sheet (GS) (curve b), caused significantly increased currents (current c).
To investigate whether the Au PNPs were necessary for increasing the electron transferring of the DNA sensors, we challenged the GCE with Au PNPs, rGO-TEPA and rGO-TEPA/Au PNPs. As seen in (Figure 2B), the signal observed in single rGO-TEPA (curve b) and Au PNPs (curve c) were significantly lower than the composite of rGO-TEPA/Au PNPs (curve d). These results clearly show that for the sensitive detection of target DNA, the presence of Au PNPs is essential. Because the Au PNPs facilitate the electron transfer of the sensor, thus amplifying the electrochemical signal.
Owing to Au PNPs can facilitate the electron transfer of the sensor, thus Au PNPs were electrodeposited on the surface of rGO-TEPA to amplify the electrochemical signal. The deposition curve is shown in (Figure 2C). Increasing the deposition time resulted in a fast increase of the current responses, followed by a flat response at approximately 30 s. This indicated that the maximum deposition of Au PNPs was obtained. To evaluate whether Au PNPs have been successfully deposited on rGO-TEPA/GCE surface, we investigated the Cyclic Voltammetry (CV) of the rGO-TEPA modified electrode and rGO-TEPA/Au PNPs modified electrode in a 0.5 M H2SO4 solution. The peaks are associated with the reduction of oxide species on the electrode surface and can be used to determine the surface composition [23]. The peak at approximately 0.9 V corresponded to Au species [24]. As shown in (Figure 2D). no peak was obvious for the rGO-TEPA modified electrode (curve a), while the rGO-TEPA/Au PNPs (curve b) generated an obvious peak at 0.9 V in a 0.5 M H2SO4 solution, demonstrating that Au PNPs had been successfully deposited on the rGO-TEPA/GCE surface. Meanwhile, using the Randles–Sevcik equation, the surface area of the rGO-TEPA/Au PNPs modified electrode is 4.81 times than that of the rGO-TEPA/GCE.
To monitor the modification procedure of the electrodes, CV experiments were performed in a 5 mM [Fe(CN)6] 3-/4- solution. As shown in (Figure 2E). A pair of typical reversible redox peaks of ferricyanide ions can be observed on the bare GCE (curve a). After the electrode was modified with rGO-TEPA, the peak current of the CV increased (curve b) due to the excellent electrical conductivity of rGO-TEPA. After Au PNPs were electrodeposited on the above electrode, the peak current further increased (curve c), implying that Au PNPs promote electron transfer. When the modified electrode was incubated with streptavidin (curve d), the peak current decreased due to the poor conductivity of streptavidin. When the modified electrode was immobilized with a biotinylated PNA probe (curve e), the peak current decreased due to the specific recognition between streptavidin and the biotinylated PNA probe, which hindered electron transfer. The peak current sharply declined after the sensor was blocked with MCH (curve f) due to the formation of a hydrophobic layer.
EIS
is one of the most powerful tools for probing the features of surface– modified
electrodes. Therefore, to further illustrate that all of the fabrication steps
were effective, the electrochemical behaviours were characterized by
electrochemical impedance spectroscopy using a 5 mM [Fe(CN)6] 3-/4- solution.
It is well-known that the semicircle diameter in EIS is equal to the electron
transfer resistance (Ret)
and that the linear part of the curve at low frequency represents the diffusion
process. As shown in (Figure 2F). when rGO-TEPA
was modified onto the electrode, the resistance decreased (curve
b) compared with the bare GCE (curve a), indicating that rGO-TEPA accelerated the electron
transfer of the electrode. When Au PNPs were electrodeposited onto the
electrode, the resistance decreased (curve c), indicating that Au PNPs promoted
electron transfer between the electrode and the solution. Afterwards, a large
semicircle diameter was observed when streptavidin was dropped onto the
electrode (curve d), implying that streptavidin was successively immobilized
onto the electrode. This larger diameter might be attributed to an insulating
layer of proteins hindering the diffusion of electrons between the electrolyte
and the electrode surface. After the electrode was incubated with the
biotinylated PNA probe (curve e), the Ret increased dramatically due to the
nonconductive property of DNA. Subsequently, with MCH blocked onto the
electrode, the resistance increased (curve f), indicating that the sensors had
indeed been successfully fabricated.
To validate that the signal amplification of our DNA assay, we investigated the dependence of the electrochemical signal on streptavidin. As expected, a significant signal change was observed in the presence of streptavidin (curve b) compared to the absence of streptavidin (curve a), shown in (Figure 3A). This results clearly show that for the sensitive detection of COX-1 DNA, the presence of streptavidin is essential. To verify the electrocatalytic current of the DNA sensors is dependent on the presence of target COX-1 DNA, we tested the modified sensors with and without target DNA. As shown in (Figure 3B). the signal monitored at rGO-TEPA/Au PNPs/GCE modified with a probe sequence complementary to a portion of the target COX-1 gene significantly increases after exposure of the electrode to a synthetic target oligonucleotide (curve c). While, for the experiment in which target DNA was not introduced to the electrode (curve b), the voltammetric peak currents dropped to the background level (curve a), indicating that the current change is dependent on target COX-1 DNA.
In
order to evaluate the catalytic efficiencies achieved with redox reporter
system, we compared the catalytic current using the redox reporter, Ru(NH3)63+,
with and without the catalytic enhancement provided by MPA and Cys (Figure 3C). shows
the typical Differential Pulse Voltammetry (DPV) measurements. No considerable
catalytic currents were observed for single Ru(NH3)63+
(curve a), however, the catalytic current for Ru(NH3)63+
is successively amplified in the presence of MPA (curve b, more than 4 times) and MPA combined Cys (curve c, more than 6 times), owing to MPA and Cys creates multiple
redox cycles for a single Ru(NH3)63+ molecule and thus amplifies the observed current.
3.4. Optimization of Experimental Conditions
To ensure the high performance of the DNA sensor, several parameters involved in the experiments, such as the concentration of rGO-TEPA, deposition time of Au PNPs, concentration of streptavidin, concentration of PNA probe, immobilization time of PNA probe and incubation time were investigated. As shown in (Figure 4A). the current response increased rapidly with increasing rGO-TEPA concentrations from 0.5 mg mL-1 to 2.0 mg mL-1. A further increase in the concentration of rGO-TEPA decreased the current change, possibly because the excessive amount of rGO-TEPA decreased the surface area of the electrode and deterred electron transfer. Thus, 2.0 mg mL-1 of rGO-TEPA was used for electrode modification. In addition, the deposition time of the Au PNPs was investigated. As shown in (Figure 4B), the current response increased with the increasing the deposition time of Au PNPs from 5 s to 30 s, and then, the current response decreased with further deposition time increases. As a result, the electrode exhibited the highest current response at a deposition time of 30 s, possibly because the excessive amount of Au PNPs decreased the surface area of the electrode and deterred electron transfer. Therefore, the optimal deposition time of the Au PNPs was 30 s.
The concentration of streptavidin is another important factor influencing the DNA sensor response. Hence, the concentration of streptavidin is shown in (Figure 4C). With an increase in the concentration of streptavidin, the current responses first increased and then started to level off. The maximum current response appeared at 100 ng mL-1. Therefore, 100 ng mL-1 was chosen as the optimal concentration of streptavidin. Furthermore, to achieve an optimal electrochemical signal, the concentration of PNA probe was systematically tested (Figure 4D). shows that the current response increased beginning at a concentration of 0.1 nM, reached a maximum value at a concentration of 10 nM, and then decreased up to a concentration of 20 nM. The results suggested that the best concentration was 10 nM. Therefore, this concentration (10 nM) was chosen for subsequent research.
To evaluate the immobilization time of PNA probe, we investigated
the current responses by varying the immobilization time of 100 pg mL-1 target DNA. As
shown in (Figure 4E). with the increase of
immobilization time, the current responses for the DNA sensor first increased
quickly and then remained steady at approximately 10 h, indicating that the
maximum immobilization of DNA sensor was obtained. To completely hybridize the target
DNA in the following step, 10 h was chosen as the optimal immobilization time. Moreover, to obtain an optimal electrochemical signal,
the incubation time was investigated, as shown in (Figure
4F). For
the reason that it would take time to reach the maximum formation of the
PNA/DNA complexes, the current responses increased with the increasing of
incubation time. As the formation of PNA/DNA complexes getting saturated, the
current responses start to level off after 60 min. Therefore, 60 min was chosen
as the optimal incubation time.
3.5. 5Analytical performance of the DNA Sensor
To evaluate the sensitivity and quantitative range of the DNA sensor under the optimal assay conditions, the performance of the DNA sensor was evaluated with different concentrations of target DNA. As indicated in (Figure 5A), the DPV peak currents increased along with increasing concentrations of target DNA in the sample solution. The sample solution containing 3 mL of the 10-fold-diluted human serum sample was used for these experiments. For quantification, we measured the DPV peak currents variation ∆I (∆I = Iafter - Ibefore, where Iafter = current after target binding and Ibefore = current before target COX-1 binding). The calibration plots showed a good linear relationship between the peak currents (∆I) and logarithmic values of the analyte concentrations over the range of 10 fM to 1 nM, as shown in (Figure 5B). The linear regression equation was ∆I (µA) =0.6827 Lg C + 0.8343 [COX-1] (M) (R2 = 0.99). The detection limit of the method was estimated to be 2.8 fM at a signal-to-noise ratio of 3σ (where σ is the standard deviation of the blank, n = 8). The lower detection limit is attributed to the streptavidin-biotin system and the ECC redox cycling. Moreover, our strategy is technically simple and obviated the use of nanomaterials or multi-enzymes for signal amplification, reducing the operation complexity and assay cost.
3.6. Specificity, Reproducibility and Stability of the DNA Sensors
To investigate whether the DNA sensors could accurately interrogate a point-mutation, we challenged the DNA sensors with other oligonucleotides, such as single-base mismatch, three-base mismatch, non-complementary sequence and blank sample (a 10-fold-diluted healthy human serum sample). The responses of the DNA sensors were measured and are shown in (Figure 5C). Athough hybridization and washing were performed at elevated procedure, compared with the current response obtained from 1 nM COX-1, the responses caused by the mismatch sequences (1 nM for each one) were weak and negligible and were as low as the blank current. This indicated the high affinity of the DNA sensors to its complementary DNA target and that the developed DNA sensors could discriminate different DNA sequences effectively. A nearly negligible current change was obtained, which indicated that the developed biosensor could potentially be used in clinical application.
To assess the reproducibility of the DNA sensors, the
fabricated electrodes were assessed by intra- and inter-assay Relative Standard
Derivations (RSDs), as shown in (Figure 5D) and (Table S2). The intra-assay precision of the
developed DNA sensors was calculated by detecting three samples containing
0.01, 1, and 100 pM of COX-1. Each sample was measured five times using five
DNA sensors prepared in parallel. For different concentrations of COX-1, the
intra-assay RSD values were less than 3.97 %. The inter-assay precision was
estimated by measuring one sample with three DNA sensors that were
independently formed at the same GCE, the RSD values were less than 4.22 % for different concentrations of COX-1.
These results indicate that the proposed DNA sensors possesses acceptable
precision and reproducibility.
To evaluate the stability of the DNA sensors, we
investigated long-term storage. The DNA sensors were stored at 4°C when it was not in use and measured periodically. As
shown in (Figure 5E), no obvious changes were
observed during the first 3 days of storage, and the current changed by less
than 1.6 %. In addition, we found that the electrode remains 97.2 % of its
initial response to 10 pM targets for 7 days of storage and only loses 18.9 %
of its sensing ability after one month of storage. Thus, the proposed DNA
sensors has satisfactory stability.
3.7. Patient-Sample Analysis
To test the potential application of the DNA sensors for
practical analysis, the DNA sensors were used to test the recovery of three
concentrations of COX-1 in human serum samples, referencing the reported
methods. Three concentrations (10 fM, 1 pM and 1 nM) of COX-1-spiked human
serum samples were prepared by the standard addition method. The serum samples
were diluted to suitable concentrations with pH 7.4 PBS (1:10). It can be
observed from (Table S3) that the relative standard deviations were in the range
of 2.249 to 3.778% and that the recoveries were in the range of 87.70 to
101.21%. These results show that the developed DNA sensors might be
preliminarily applied to determine the presence of COX-1 in real samples.
The ultimate goal of developing a mutation-discriminating DNA sensor was to enable the direct analysis of mutated sequences in patient samples. We challenged our DNA sensors with the task of analysing COX-1 gene in serum samples from four patients who were treated invalidly with aspirin. Ten microliter samples of the four different patients were used. Each human serum sample was measured 5 times. As illustrated in (Figure 5F), the DNA sensors showed different current response for each of the patient serum sample. From the current response, the concentrations of COX-1 in the four different invalidly treated patients with aspirin were calculated to be 23.7 fM (Sample 1, Patient 1), 96.2 fM (Sample 2, Patient 1), 54.5 fM (Sample 3, Patient 3) and 272.8 fM (Sample 4, Patient 4), respectively. To ascertain the correctness of the results, we also tested the COX-1 contents with qRT-PCR. The results shown in Fig. 5F indicated that the levels of COX-1 in three real samples obtained by our method and qRT-PCR were of the same order of magnitude. Thus, the results indicated that the proposed DNA sensors could easily determine different concentrations of COX-1 and showed a high sensitivity.
4. Conclusion
In this paper, a novel and ultrasensitive electrochemical DNA sensor for the detection of COX-1 was prepared using ECC redox cycling. The proposed method has the following advantages: (1) a neutral charge biotinylated Peptide Nucleic Acid (PNA) probe modify the electrode surface can minimize the background current and increase the signal-to-noise ratio; (2) the newly introduced chemical regents of MPA and cysteamine not only permit Ru (NH)3+ to be regenerated for multiple redox cycles but also overcome the interfering redox reactions near the potential of interest; (3) the detection limit is low. The low detection limit indicates that this approach is applicable to clinical pre-administration detection. Accordingly, this methodology could simplify and speed up the identification of drug-susceptible people, saving medical resources and reducing time to treatment. Although we have focused on the detection of COX-1 mutations for aspirin, the DNA sensor can be extended to other relevant mutations recurring in other types of medicine-related gene. So, the presence of multiple mutations may be interrogated simultaneously, allowing a more detailed and precision clinical administration.
5. Acknowledgements
The
research was financed by the Science and technology project of Sichuan
provincial health and Family Planning Commission (No. 16PJ550).
Scheme 1: Schematic illustration of the electrochemical DNA
bioassay protocol.
Figure 1(A-D): SEM images of (A)
rGO-TEPA, (B) rGO-TEPA/Au PNPs, (C) rGO-TEPA/AuPNPs/streptavidin, and (D) rGO-TEPA/Au PNPs/streptavidin/PNA
probe. The inset of (a) shows the FE-SEM image of rGO-TEPA; The inset of (b) shows
the TEM image of Au PNPs.
Figure 2(A-F): (A) The
CVs of (a) bare GCE, (b) GS/GCE, and (c) rGO-TEPA/GCE in a 5 mM [Fe(CN)6]
3-/4- solution containing a 0.1 M KCl solution. (B) DPV responses of the DNA sensor of
different modification steps: (a) bare GCE; (b) GCE/rGO-TEPA; (c) GCE/Au PNPs; and
(d) GCE/ rGO-TEPA /Au PNPs in 5 mM [Fe(CN)6] 3-/4- containing a 0.1 M KCl solution. (C) The deposition curve of Au PNPs. (D) CV curves of GCE/rGO-TEPA before (a) and after (b) deposition of porous
gold nanoparticles in 0.5 M H2SO4 from 0 to 1.5 V at a scan
rate of 100 mV/s. (E) CV
characterization of electrodes at various stages of modification in a 5 mM
[Fe(CN)6] 3-/4- solution
containing 0.1 M KCl: (a) bare GCE, (b) GCE/rGO-TEPA, (c) GCE/rGO-TEPA/Au PNPs (d) GCE/rGO-TEPA/Au PNPs/streptavidin, (e) GCE/rGO-TEPA/Au PNPs/streptavidin/PNA probe, (f) GCE/rGO-TEP/AuNPs/AuPNPs/streptavidin/PNA probe/MCH; (F) EIS for each immobilization step in 5 mM [Fe(CN)6] 3-/4-
containing a 0.1 M KCl solution: (a) bare GCE, (b) GCE/rGO-TEPA, (c) GCE/rGO-TEPA/Au PNPs (d) GCE/rGO-TEPA/Au
PNPs/streptavidin, (e) GCE/rGO-TEPA/Au PNPs/streptavidin/PNA probe, (f)GCE/rGO-TEPA/Au PNPs/streptavidin/PNA probe/MCH in 5 mM [Fe(CN)6]3-/4- containing a 0.1 M KCl solution.
Figure 3(A-C): (A) Cyclic
voltammograms of the electrodes without (curve a) and with (curve b)
streptavidin modification in 5 mM [Fe(CN)6]3-/4- containing
a 0.1 M KCl solution. (B) Cyclic
voltammograms of the modified electrode before (curve b) and after (curve c)
hybridization with target DNA in 5 mM [Fe(CN)6]3-/4- containing
a 0.1 M KCl solution. (C) Different Pulse
Voltammetry (DPV) scans of DNA sensors modified with PNA/DNA duplexes and incubated in 10 µM Ru(NH3)6Cl3 for 60 s and then measured in a
solution of 50 mM Tris buffer solution pH 9 (curve a), containing 0.5 mM 3-Mercaptopropionoic
Acid (MPA) (curve b) as catalytic enhancer, and containing 0.5 mM
3-mercaptopropionoic acid (MPA) and 0.5 mM cysteamine (curve C) as catalytic
enhancer.
Figure 4(A-F): Effects of (A)
the concentration of rGO-TEPA, where n = 10 for each point. (B) deposition time of Au PNPs, where n
= 10 for each point. (C)
concentration of streptavidin and (D)
PNA probe, where n = 10 for each point. (E)immobilization
time of PNA probe on the modified electrode, and (F) incubation time, where n = 10 for each point.
Figure 5(A-F): (A) DPV signals of the DNA sensors in the presence of different concentrations of COX-1
and (B) the calibration curve of the DNA sensors for 10 fM, 100 fM, 1 pM, 10 pM, 100 pM, 1 nM of COX-1 (n = 3). (C) Specific response of the DNA
sensors hybridized with complementary COX-1, single-base mismatch, three-base
mismatch, non-complementary and blank sample (100 pM for each analyse). (D) Reproducibility of 7 different
electrodes modified with 100 pM of VKORC1. (E)
The stability of the COX-1 DNA sensors (100 pM) after 3 d, 7d, 10 d, 14 d, 21
d, and 28 d. (F) Electrochemical currents of COX-1 in four serum samples detected by the presented method
(black bars) and qRT-PCR (red bars), respectively.
Nucleotide name |
Sequence (5’®3’) |
PNA |
Bio-Cys-O-ATC-ATC-CTG-ACT-GGC-ATC-CGG |
Target DNA |
CCG-GAT-GCC-AGT-CAG-GAT-GAT |
Single-base mismatch |
CCG-GAT-GGC-AGT-CAG-GAT-GAT |
Three-base mismatch |
CCG-GAT-GGC-AGT-CTG-GTT-GAT |
Non-complementary |
GGC-TGA-GGC-TAC-GGC-ACT-G |
Table S1: Synthetic oligonucleotide sequences in this study.
COX-1 (pM) |
|
0.01 |
1 |
100 |
RSD (%) |
Intra-assay |
3.97 |
3.22 |
3.03 |
|
Inter-assay |
4.07 |
4.22 |
2..95 |
Table S2: Reproducibility of the electrochemical detection of COX-1.
Samples |
Added |
Found* |
Recovery |
RSD |
|
|
|
(%) |
(%) |
Sample-1 |
0.01 pM |
0.877 pM |
87.7 |
3.778 |
Sample-2 |
01:00 PM |
0.998 pM |
99.8 |
3.012 |
Sample-3 |
100 pM |
101.21 pM |
101.21 |
2.249 |
Samples (1-3) were from healthy people. |
||||
*The values shown here are the average values from five measurements. |
5.
Nojoomi F, Vafaei M, Bagheri Z, Zamanian Z (2017) Evaluation
of Aspirin Effect on Candida Glabrata Isolates. MOJ Immunol 5: 00173.
11.
Zhao Y, Hu S, Wang H, Yu K, Guan Y, Liu X, Li N, Liu F, Analytical Chemistry, 2017, Epub
ahead of print.