Performance Characteristics of Spect Imaging System in Clinical Environment
Saeed-Ur-Rahman1*, Maria Alvi2, Syed Ali Mehdi2, Shazia Fatima1, Mohammad Faheem1
1Department of Nuclear Medicine, Oncology and
Radiotherapy Institute (NORI), Islamabad, Pakistan
2Department of Physics, International Islamic University, Islamabad, Pakistan
*Corresponding author: Saeed-Ur-Rahman, Department of Nuclear Medicine, Oncology and Radiotherapy Institute (NORI), Islamabad, Pakistan. Tel: +923008309821; Email: snori66@gmail.com
Received Date: 01 July, 2017; Accepted Date: 05 July, 2017; Published Date: 16 July, 2017
Citation: de la Cruz A, Logsdon R, Lye D, Guglielmi S, Rice A, et al. (2017) Harmful Algae Bloom Occurrence in Urban Ponds: Relationship of Toxin Levels with Cell Density and Species Composition. J Earth Environ Sci: JEES-148. DOI: 10.29011/ JEES-148. 100048
1. Abstract
Quality control has a fundamental role in the nuclear medicine before performing any medical examination to ensure the particular aspects of the procedure and to compliance the regulatory requirement. Optimizing the imaging equipment before clinical use, involves the measurement of radioactivity, selection of proper instrumentation, reconstruction technique and to perform the imaging study properly. The aim of this study is to assess the performance characteristics of dual head SPECT gamma camera installed at the nuclear medicine department of NORI, Islamabad.
Different methodologies have been developed by the manufacturer and various international organizations to evaluate the system performance. Radioisotopes involving Tc-99m, Co-57 and I-131 of different amount are used according to respective calibration measurements. Tests performed during studies include the measurement of system differential and integral uniformity, energy resolution, image resolution and linearity, center of rotation and sensitivity.
Results obtained by these
experimental studies showed uniformity and energy resolution values are in the
range and within limits recommended by IEC and AAPM. Center of rotation for
each detector configuration is measured and found acceptable. Intrinsic
resolution is found better compared to extrinsic resolution, sensitivity values
were sustainable and good linearity behavior of gamma camera observed.
Unsealed radioactive sources are used for diagnostic as well as therapeutic purpose to perform various kinds of procedures in nuclear medicine. The most commonly used radioisotopes in nuclear medicine at NORI are 99mTc and 131I having half life of 6 hrs and 8 days, respectively. Precision and accuracy in diagnostic studies can be achieved by understanding the standards of the techniques used for maintaining and optimizing the performance of the devices being used. The purpose of the present was to perform the acceptance testing of Siemens Symbia dual head gamma camera newly installed at NORI in order to evaluate its performance characteristics for clinical use. The research study was conducted to measure its performance on the basis of daily, weekly and monthly quality control tests. The recommended tests were carried out by applying IEC, NEMA and other international organizations protocols to ensure the scintillation camera is performing accurately to the specifications that are provided by the manufacture and compliance with the regulatory bodies [1-4].
The performance
evaluation of the gamma camera was carried out in accordance to the recommended
procedures and techniques of different professional organizations. In addition
to these guidelines different protocols and workflows are locally developed to
assess the performance of the system before using it clinically. This study is
based upon not only on the test performance and the evaluation of results
according to given protocols and recommendations but to fulfill the standards
and results required on daily basis.
The tests performed during study on the system are energy peaking of the detector, flood field uniformity, resolution of the system, center of rotation and total SPECT performance. In addition to these procedures, regular QC processes, periodic tests are also carried out to evaluate in-depth details and performance characteristics of the system. All these tests provide data against system subsequent performance on weekly, monthly, quarterly and annually basis. Operational checks are carried out on daily basis to check performance of the instruments and detail records are maintained. The maintained records are useful for the comparative study and future reference [5-8].
Radioactive sources used to analyze the performance characteristics of SPECT system on the basis of performed tests were flood sheet source of 57Cobalt, point sources of technetium and iodine radioisotopes. Activity of the respective source is measured with dose calibrator and poured on a small piece of cotton approximately 3mm which is placed in the vial. It is essential for the point source preparation that the specific activity of the source must be high enough otherwise it may affect the results of gamma camera.
Different types of collimators are used for the extrinsic measurement studies of given Gamma Camera. Collimators common properties includes its thickness, maximum energy rating and it is hole angulations (diverging/converging hole). Quadrant bar phantom is used for measuring spatial resolution and distortion testing. It is for accurate measurements of camera intrinsic resolution, collimators spatial resolution, field size and linearity. Jaszczak SPECT phantom is used for the evaluation of total SPECT performance of the system. It is designed to evaluate the overall SPECT performance. The phantom is mainly useful for identifying whole degradation of the SPECT system. It measures the system resolution contrast and uniformity at the same time. A series of spheres and rods of variable diameters are present in the phantom [9-13].
Different source holders were used in performing the recommended tests. These source holders include, five points source holder, sheet source holder and integrated source holder.
Five points source holder is designed for point sources used to perform different tests especially for the performance of COR and is also known as COR phantom. Whereas sheet source holder is specially designed for flood sheet source and is used mostly in extrinsic studies. Another Integrated source holder which is attached to the rear bed pallet located at the back of the gantry [10,11].
Acceptance tests of the SPECT system, Siemens Symbia Gamma Camera which is newly installed at NORI are performed and evaluated. Detail of the performed tests and corresponding results are given in this section.
Pulse
height analyzer from the processing panel was used to perform the energy
peaking of the system and radioactive 57Co
source of 10 m Ci is used to evaluate the energy resolution of the system. Results
obtained for energy peaking, window width and dead time for both detectors is
given in Table 1.
It is observed from the obtained results that energy window width for each detector is estimated about 20%. The measured peak shifts values obtained in the present study are 1.35 and 1.27 for detectors 1 and 2, respectively. The peak shift values are within the recommended limits (+3.0%). Dead time measured for detectors 1 and 2 found to be 2.26 and 5.00, respectively. These dead time values lie within the recommended range as given by the NEMA protocols.
Flood
field uniformity evaluation includes both the integral and differential
uniformity over central and useful fields of view. Results obtained for both
detectors are shown in table 2. Data given in tables 2 showing calculated
values of intrinsic and extrinsic uniformity calibration. Integral uniformity
for detector 1 with CFOV and UFOV are 3.98% and 4.36 % respectively for
intrinsic uniformity. Similarly, for differential uniformity, the results
obtained for detector 1 are 1.80% and 2.37%. The results obtained for detector
2 having integral and differential uniformity for CFOV and UFOV found to be 4.09
%, 4.75% and 2.10%, 2.53% respectively. Extrinsic uniformity results for both
detectors are also elaborated in Table 2.
The calculated results for both detectors showing integral and differential values in table 2 for intrinsic and extrinsic uniformities are within range and acceptable limits. Differential uniformity values are less than the integral values as expected because differential uniformity measures the variation over the small field whereas integral uniformity is the measure of variation over the entire detector.
The Center of Rotation (COR) was performed for the verification of SPECT performance as well as for the mathematical formalism of tomographic image reconstruction. The study was performed for Low Energy High Resolution (LEHR) collimator which is commonly used for clinical studies. The COR test includes all detector configuration e.g. 76°, 90° 180° and acquired data displays a sinogram for visual inspection and calculated values for COR deviations. Tabulated form of the obtained results is given below in the Table 3.
Results obtained for both detectors configuration are acceptable and lies within the limits. System resolution is depending on the collimator used and obtained results for both of the detectors show COR values are acceptable. Regarding these results the system working especially for SPECT studies is considered normal and sustaining.
A
common approach for intrinsic and system resolution is the visual evaluation of
the bar phantom images. The minimum distinguishable bar spacing in the acquired
static phantom images is used as an index of camera spatial resolution. A
quadrant bar phantom is used for performing the test to find out the resolution
of the system. Images are acquired for both detectors as well as for intrinsic
and extrinsic resolution with different angles. Images acquired with the bar
phantom by rotating it 900 for each
study to assess the linearity of both detectors are shown in figure 1. The
acquired images show a resolution of the system which is acceptable and within
the recommended limits
Acquired images
for intrinsic and extrinsic resolution are showing that intrinsic resolution
images are better than the extrinsic images. Minimum intrinsic resolution is
2.5 mm whereas extrinsic resolution is 3 mm according to minimum distinct bar
spacing on the phantom.
The
test used for the assessment of overall SPECT performance in a single scan
using Jaszczak phantom. This phantom has special design having spheres and rods
that evaluates the resolution uniformity and its contrast at the same time. The
results for Jaszczak phantom are shown in Figure 2.
Visual
evaluation of the obtained images shows that the area without spheres are
uniform and no ring arte facts or any kind of non-uniformities are observed
here. Cold area of spheres inside the phantom for system contrast is properly
distinguishable and showing sustainable contrast for clinical studies. For
system resolution out of six, five spheres are visible with the minimum
resolution of 12.7mm. These images are consistent with the images obtained at
the time of acceptance. So, it can be concluded from this study that system is
optimum for the proper clinical performance.
4. Conclusion
The study was focused to analyze
the performance of a newly installed SPECT system at NORI by carrying out
acceptance testing before getting into its routine clinical use. Performance of
the system was specified in terms of its energy peaking, uniformity, center of
rotation, resolution and linearity. Results obtained for energy peaking are
acceptable as the measured peak shifts and dead time are within the recommended
limits. System uniformity is within the recommended limits on the basis of calculated
values. The results for center of rotation, axial shift, and back projection
angle are within acceptable range for which the SPECT system working is
considered sustaining. System resolution and linearity are evaluated visually
which shows that intrinsic resolution is little better than extrinsic
resolution whereas system linearity is showing good behavior which is
sustainable for clinical use. The protocols and workflows developed locally are
in routine use. Hence the system is accepted for clinical use on the basis of
obtained results.
Figure 1: Intrinsic and extrinsic resolution of
detectors 1 and 2.
Figure 2: Jaszczak
phantom images.
Detector1 |
Detector2 |
|
Width % |
20 |
20 |
Peak shift % |
1.35 |
1.27 |
Peak status |
Peaked |
Peaked |
Dead time % |
2.26 |
5 |
Table 1: Energy peaking and dead time for Detectors 1 and 2.
|
Intrinsic |
Extrinsic |
|||||
Detector |
Uniformity |
Central field of view |
Useful field of view |
Central field of view |
Useful field of view |
||
1 |
Integral |
3.98% |
4.36% |
3.22% |
3.65% |
||
|
Differential |
1.80% |
2.37% |
2.25% |
2.31% |
||
2 |
Integral |
4.09% |
4.75% |
3.10% |
3.92% |
||
|
Differential |
2.10% |
2.53% |
2.03% |
2.77% |
Table 2: Integral and differential uniformities for detectors 1 and 2.
Parameters |
COR at 760 |
COR at 900 |
|
COR at 1800 |
|
|||||
Detector 1 |
Detector 2 |
Detector 1 |
|
Detector 2 |
Detector 1 |
Detector 2 |
||||
Center of rotation |
-0.526 mm |
-2.069 mm |
-0.661 mm |
|
-2.145 mm |
-0.479 mm |
-1.925 mm |
|||
|
|
|
|
|
|
|
|
|||
Axial shift |
-0.679mm |
0.679 mm |
-0.756 mm |
|
0.756 mm |
-0.493mm |
0.493 mm |
|||
Back projection angle |
-0.0410 |
0.0410 |
-0.0220 |
|
0.0220 |
-0.0560 |
0.0560 |
|||
System resolution at 20 cm |
18.008 mm |
17.933 mm |
22.431 mm |
|
22.332 mm |
14.667 mm |
14.620 mm |
Table 3: COR of the SPECT system at different angles.
- IEC-60789 (2005) Medical electrical
equipment, Characteristics and test conditions of radionuclide imaging devices
- Anger type gamma cameras.
- IAEA Human Health Series No
(2006) 6 International Atomic Energy Agency, Quality Assurance for SPECT
Systems Vienna.
- Siemens Operating Instructions Symbia System (2010) Quality Control and Assurance, T16/T6/T2/T Series, Version VA60A.
- NEMA Standards Publication NU (1-2012) National Electrical
Manufacturers Association, Performance Measurements of Gamma Cameras.
- Zanzonico P (2008) Routine quality control of clinical
nuclear medicine instrumentation. A brief review. J Nucl Med 49: 1114-1131.
- Busemann SE, Płachcinska A, Britten A (2010) Routine quality control recommendations for nuclear medicine instrumentation. Eur J Nucl Med Mol Imaging 37: 662-671.
- Bolstad R, Brown J, Grantham V (2011) Extrinsic Versus
Intrinsic Uniformity Correction for gamma-cameras. J Nucl Med Technol 39: 208-212.
- Bhatia B.S, Bugby SL, Lees JE, Perkins AC (2015) A scheme for assessing the performance characteristics of small field-of-view gamma cameras. Physica Medica 31: 98-103.
- Bailey DL, Aswegen A, Van A, Humm JL, Todd PA (2014) International Atomic Energy Agency IAEA, Division of Human Health, Vienna Austria: 312-397.
- Nichols KJ, Bacharach SL, Bergmann SR, Cullom SJ, Ficaro EP, et al. (2007) Links J, Inst quality Ass and perf. J Nucl Cardio 14: 61-78.
- Muehllehner G, Wake RH, Sano R (1981) Standards for performance measurements in scintillation cameras. J Nucl Med 22: 72-77.
- Murphy PH (1987) Acceptance testing and quality control of
gamma cameras, including SPECT. J Nucl Med 28: 1221-1227.
- Alehyani SHA (2009) Application of single photon emission computed tomography (SPECT) parameters for bone scintigraphy. J King Saud Univ Sci 21: 109-117.